Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 601

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 653
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 600
SCPI-Befehlsreferenz
OSA 593
Product-Specific Commands—Description
:TRACe[:DATA]:X:STOP[:WAVElength]?
Description This query returns the X magnitude of the last
point for the specified trace.
This query has no associated *RST condition.
Syntax :TRACe[:DATA]:X:STOP[:WAVElength]?<wsp><
TraceName>
Parameter(s) TraceName:
The program data syntax for <TraceName> is
defined as a <STRING PROGRAM DATA>
element.
The <TraceName> parameter corresponds to
the name of the trace to select.
Response Syntax <Stop>
Response(s) Stop:
The response data syntax for <Stop> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Stop> response corresponds to the X-axis
wavelength of the last point of the trace
expressed in meter.
Przeglądanie stron 600
1 2 ... 596 597 598 599 600 601 602 603 604 605 606 ... 652 653

Komentarze do niniejszej Instrukcji

Brak uwag