SCPI-Befehlsreferenz
542 FTB-5240S/S-P/BP
Product-Specific Commands—Description
SENS:AVER:COUN 8
SENS:WAV:STAR 1525.000 NM
SENS:WAV:STOP 1570.000 NM
TRIG:SEQ:SOUR IMM
STAT? Poll until returned state is READY
INIT:IMM
STAT:OPER:BIT8:COND? Poll until returned state
is 0
Notes The trigger system leaves IDLE state to perform
acquisition only if the instrument is in READY
status.
This command is used to start single, averaging,
InBand or i-InBand acquisitions.
See Also :ABORt
:CALCulate[1..n][:WDM]:STATe
:INITiate:CONTinuous
:SENSe[1..n]:AVERage:STATe?
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?
:TRACe:FEED:CONTrol
:TRIGger[1..n][:SEQuence]:SOURce
:INITiate[:IMMediate]