SCPI-Befehlsreferenz
OSA 521
Product-Specific Commands—Description
Response(s) Resolution:
The response data syntax for <Resolution> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Resolution> response corresponds to
either the current or the MINimum/MAXimum
resolution bandwidth value for OSNR calculation.
Example(s) CALC:WDM:OSNR:BAND:RES:AUTO OFF
CALC:WDM:OSNR:BAND:RES 0.065 NM
CALC:WDM:OSNR:BAND:RES? Returns
6.500000E-011
See Also :CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDw
idth[:RESolution]:AUTO
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDw
idth[:RESolution]
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]?