Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 515

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 653
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 514
SCPI-Befehlsreferenz
OSA 507
Product-Specific Commands—Description
The <Select> response corresponds to the
name of the selected channel result. A single null
string is returned if no channel has been
selected.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:DEF "C_001", 1300.000 NM
<Do measurement>
CALC:WDM:DATA:CHAN:SEL? Returns "" (no
selection)
CALC:WDM:DATA:CHAN:SEL "C_001"
CALC:WDM:DATA:CHAN:SEL? Returns "C_001"
See Also :CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:CATalo
g?
:CALCulate[1..n][:WDM]:DATA:CHANnel:NSELec
t?
:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:
SELect?