Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 533

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 653
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 532
SCPI-Befehlsreferenz
OSA 525
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:STATe
Description This command controls activation of WDM
analysis.
At *RST, WDM analysis is set to on (enabled).
Syntax :CALCulate[1..n][:WDM]:STATe<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of WDM analysis.
0 or OFF: WDM analysis is disabled.
1 or ON: WDM analysis is enabled.
Example(s) CALC:WDM:STAT ON
CALC:WDM:STAT? Returns 1 (WDM analysis
enabled)
Notes WDM analysis cannot be disabled: OFF (0) value
is valid for queries only.
It is possible to query acquired trace data only if
active measurement analysis mode is WDM.
See Also :CALCulate[1..n][:WDM]:STATe?
:INITiate[:IMMediate]
:INITiate:CONTinuous
:TRACe:FEED:CONTrol
Przeglądanie stron 532
1 2 ... 528 529 530 531 532 533 534 535 536 537 538 ... 652 653

Komentarze do niniejszej Instrukcji

Brak uwag