P/N: 1066920 July 2014 1/6
NOTICE
This information applies only to the
power meter and VFL designed for the
FTB-2 and FTB-2 Pro units.
Installing or Removing the Power Meter and VFL
If your unit was not equipped with a power meter and a VFL at time of your
initial purchase, you can purchase one afterwards and install it yourself.
You may also need to remove the power meter to send it to an authorized
service center for recalibration.
Note: If you prefer not to remove the power meter from your unit when a
calibration is needed, you can also send the complete unit to the service
center.
WARNING
To avoid serious injuries as well as irreparable damages both to
your unit and your power meter, ALWAYS TURN OFF YOUR UNIT
(shutdown mode, not sleep), DISCONNECT IT FROM THE
EXTERNAL POWER SOURCE, and REMOVE ALL BATTERIES. For
more information on the ways to turn off your unit, refer to the
user guide or the online help of the FTB-2 and FTB-2 Pro.
Be careful not to drop metal objects such as screws inside the
unit. These could cause a short-circuit leading to a fire or an
explosion.
Przeglądanie stron 0
1 2 3 4 5 6

Podsumowanie treści

Strona 1 - FTB-2 and FTB-2 Pro units

P/N: 1066920 July 2014 1/6NOTICEThis information applies only to thepower meter and VFL designed for the FTB-2 and FTB-2 Pro units.Installing or Remov

Strona 2 - P/N: 1066920 July 2014 2/6

P/N: 1066920 July 2014 2/6CAUTIONElectrostatic discharge (ESD) damage can cause complete or intermittent equipment failures.  Always use an ESD-preve

Strona 3

P/N: 1066920 July 2014 3/6To install a power meter:1. Turn off your unit.2. Disconnect the unit from its external power source and remove the batterie

Strona 4 - Writings

P/N: 1066920 July 2014 4/67. By holding the power meter by its faceplate, position it so that the writings are right-side up.8. Carefully align the po

Strona 5

P/N: 1066920 July 2014 5/6To remove the power meter for recalibration:1. Turn off your unit.2. Disconnect the unit from its external power source and

Strona 6 - P/N: 1066920 July 2014 6/6

P/N: 1066920 July 2014 6/66. By holding the power meter by its faceplate, gently pull it out of the bay completely.7. Place the power meter in an anti