SCPI-Befehlsreferenz
OSA 527
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:THReshold
Description This command sets WDM analysis absolute
power threshold for peak detection.
At *RST, this value is set to -45.0 dBm.
Syntax :CALCulate[1..n][:WDM]:THReshold<wsp><Th
reshold[<wsp>DBM|W]>|MAXimum|MINimu
m|DEFault
Parameter(s) Threshold:
The program data syntax for <Threshold> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: DBM|W. The
<Threshold> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.