Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 384

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 653
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 383
SCPI-Befehlsreferenz
376 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:AUTO?
Description This query indicates if i-InBand auto noise
measurement for WDM analysis of the default
channel has been enabled or not.
At *RST, auto noise measurement is set to off
(disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO?
Parameter(s) None
Response Syntax <Auto>
Response(s) Auto:
The response data syntax for <Auto> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Auto> response corresponds to the state
of the auto noise measurement.
0: auto noise measurement is disabled.
1: auto noise measurement is enabled.
Example(s) CALC:CHAN:AUTO:NOIS:AUTO OFF
CALC:CHAN:AUTO:NOIS:AUTO? Returns 0 (auto
noise disabled)
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO
Przeglądanie stron 383
1 2 ... 379 380 381 382 383 384 385 386 387 388 389 ... 652 653

Komentarze do niniejszej Instrukcji

Brak uwag