Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 633

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 688
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 632
Справка по командам SCPI
OSA 627
Product-Specific Commands—Description
The <Data> response corresponds to the Y-axis
values vector of the trace. Returned values are
evenly spaced relative to the X-axis expressed in
meter.
X-axis wavelength interval between each Y value
is determined as follow:
interval = (stop - start) / (count - 1) where:
start = TRACe[:DATA]:X:STARt[:WAVelength]?
stop = TRACe[:DATA]:X:STOP[:WAVelength]?
count = TRACe:POINts?
The points unit is determined by the trace
definition context. When trace data represents
absolute power, returned values are in dBm.
When trace data represents relative power,
returned values are in dB.
Example(s) <Do measurement>
FORMat:DATA ASC
TRAC:DATA? "TRC1" Returns
-5.246202E+001,-5.246195E+001,-5.246181E+0
01,...
FORMat:DATA PACK
TRAC:DATA? "TRC1" Returns binary data
:TRACe[:DATA][:Y][:WAVElength]?
Przeglądanie stron 632
1 2 ... 628 629 630 631 632 633 634 635 636 637 638 ... 687 688

Komentarze do niniejszej Instrukcji

Brak uwag