Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 521

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 688
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 520
Справка по командам SCPI
OSA 515
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:DATA:CHANnel:
CenterMASs:FREQuency?
Description This query returns computed WDM analysis result
for center of mass frequency of the selected
channel.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:DATA:CHANnel:Center
MASs:FREQuency?
Parameter(s) None
Response Syntax <Center>
Response(s) Center:
The response data syntax for <Center> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Center> response corresponds to the
computed center of mass frequency in hertz.
Example(s) CALC:CHAN:DEL:ALL
CALC:CHAN:DEF "ITU_14", 201.9873 THZ
<Do measurement>
CALC:DATA:CHAN:SEL "ITU_14"
CALC:DATA:CHAN:CMAS:FREQ? Returns
2.020066E+014
Przeglądanie stron 520
1 2 ... 516 517 518 519 520 521 522 523 524 525 526 ... 687 688

Komentarze do niniejszej Instrukcji

Brak uwag