Справка по командам SCPI
OSA 579
Product-Specific Commands—Description
:MEMory:TABLe:SELect
Description This command selects the table to define.
At *RST, there is no selection: a single null string
is returned.
Syntax :MEMory:TABLe:SELect<wsp><TableName>
Parameter(s) TableName:
The program data syntax for <TableName> is
defined as a <STRING PROGRAM DATA>
element.
The <TableName> parameter corresponds to the
name of the table to select.
Example(s) <Do measurement>
MEM:TABL:SEL "WDM:CHANNEL"
MEM:TABL:DEF "NAME,CMAS:WAV"
MEM:TABL:POIN? "WDM:CHANNEL" returns 2
MEM:TABL:DATA? "WDM:CHANNEL" returns
#248"C_001,1.55236113E-006","C_002,1.55672
7357E-006"
Notes The only valid table name is "WDM:Channel".
See Also :MEMory:TABLe:DEFine
:MEMory:TABLe:DEFine?