Справка по командам SCPI
620 FTB-5240S/S-P/BP
Product-Specific Commands—Description
The <Condition> response represents the current
operation condition of the instrument. The
meaning of the response depends on the value
returned for bit <n>.
bit <8>: When the returned value is 1, the
instrument is performing a measurement (trigger
system INITiated).
bit <9>: When the returned value is 1, the
instrument is performing an offset nulling and/or a
wavelength referencing
(CALibration:ZERO:AUTO?).
Example(s) STAT? Must return READY
CAL:ZERO:AUTO ONCE
STAT:OPER:BIT9:COND? Keep resending this
query as long as the operation is not complete
(returned value is not 0).
See Also :CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus?
:STATus:OPERation:BIT<8|9>:
CONDition?
Przeglądanie stron 625