Справка по командам SCPI
OSA 485
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Width> parameter.
The <Width> parameter corresponds to a valid
width in meter for the noise measurement region.
The
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh[
:WAVelength]? MIN and
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh[
:WAVelength]? MAX queries can be used to
determine valid width values.
Example(s) CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ
CALC:WDM:CHAN:SEL "ITU_22"
CALC:WDM:CHAN:NOIS:TYPE POLY5
CALC:WDM:CHAN:NOIS:WIDTh:WAV 12.5 NM
CALC:WDM:CHAN:NOIS:WIDTh:WAV? Returns
1.250000E-008
Notes Custom width for noise measurement region is
applied only if selected noise type is
POLYnomial5.
See Also :CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh
:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh
[:WAVelength]?
:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTa
nce[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE
:CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:WIDTh[:WAVelength]