Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Instrukcja Użytkownika Strona 575

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 659
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 574
Referencia de instrucciones SCPI
OSA 569
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:OFFSet
[:MAGNitude]?
Description This query returns a value indicating either the
current or the minimum/maximum power offset.
At *RST, this value is set to 0.0 dB.
Syntax :SENSe[1..n]:CORRection:OFFSet[:MAGNitude]?[
<wsp>MAXimum|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Offset>
Przeglądanie stron 574
1 2 ... 570 571 572 573 574 575 576 577 578 579 580 ... 658 659

Komentarze do niniejszej Instrukcji

Brak uwag