Referencia de instrucciones SCPI
508 FTB-5240S/S-P/BP
Product-Specific Commands—Description
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
The <Select> parameter corresponds to a valid
channel result index to select. The channel index
cannot be zero.
The
CALCulate[1..n][:WDM]:DATA:CHANnel:COUNt?
query can be used to determine valid index
range.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:DEF "C_007", 1380.000 NM
<Do measurement>
CALC:WDM:DATA:CHAN:NSEL 1
See Also :CALCulate[1..n][:WDM]:CHANnel:NSELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:COUNt?
:CALCulate[1..n][:WDM]:DATA:CHANnel:NSELec
t?
:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:
NSELect