Referencia de instrucciones SCPI
OSA 559
Product-Specific Commands—Description
Example(s) SENS:AVER:STAT ON
SENS:AVER:TYPE SCAL
SENS:AVER:COUN? MIN Returns 2
SENS:AVER:COUN? MAX Returns 9999
SENS:AVER:COUN 20
SENS:AVER:COUN? Returns 20
Notes If averaging type is set to PMMH and auto noise
measurement is active then specific averaging
count setting has no effect. It is automatically
determined by the instrument.
See Also :INITiate[:IMMediate]
:INITiate:CONTinuous
:SENSe[1..n]:AVERage:STATe
:SENSe[1..n]:AVERage:TYPE
:SENSe[1..n]:AVERage:COUNt?