Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Instrukcja Użytkownika Strona 385

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 659
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 384
Referencia de instrucciones SCPI
OSA 379
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:AUTO
Description This command controls activation of i-InBand
noise measurement for WDM analysis default
channel.
At *RST, auto noise is set to off (disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
AUTO<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.
The <Auto> parameter corresponds to the new
state of auto noise measurement.
0 or OFF: disable default channel auto noise.
1 or ON: enable default channel auto noise.
Przeglądanie stron 384
1 2 ... 380 381 382 383 384 385 386 387 388 389 390 ... 658 659

Komentarze do niniejszej Instrukcji

Brak uwag