Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Instrukcja Użytkownika Strona 464

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 659
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 463
Referencia de instrucciones SCPI
458 FTB-5240S/S-P/BP
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Width> parameter.
The <Width> parameter corresponds to a valid
width in meter for the noise measurement
region.
The
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh
[:WAVelength]? MIN and
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh
[:WAVelength]? MAX queries can be used to
determine valid width values.
Example(s) CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ
CALC:WDM:CHAN:SEL "ITU_22"
CALC:WDM:CHAN:NOIS:TYPE POLY5
CALC:WDM:CHAN:NOIS:WIDTh:WAV 12.5 NM
CALC:WDM:CHAN:NOIS:WIDTh:WAV? Returns
1.250000E-008
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:WIDTh[:WAVelength]
Przeglądanie stron 463
1 2 ... 459 460 461 462 463 464 465 466 467 468 469 ... 658 659

Komentarze do niniejszej Instrukcji

Brak uwag