Exfo PM-1600 High-Speed Power Meter Instrukcja Użytkownika Strona 149

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 148
Remote Control (SCPI) Commands
High-Speed Power Meter 139
Specific Commands
SENSe:POWer:WAVElength
Description This command selects a new operating wavelength (for the
selected channel when using a PM-1620). If this command is
sent while an acquisition is in progress, the “Acquisition already
running” error message will be raised.
Syntax SENS:POW:WAVE<space><numeric_value> [<space>NM]
Parameters The <numeric_value> parameter is an operating wavelength
in nm. Any wavelength within the spectral range of the power
meter optical detector at 0.01 nm resolution may be selected.
See the power meter instruction manual for the exact spectral
range of each detector type. An invalid parameter will raise a
“Data type error” message.
Example SENS:POW:WAVE 1310.12
SENSe:POWer:WAVElength?
Description This query returns the currently selected wavelength (for the
selected channel when using a PM-1620).
Syntax SENS:POW:WAVE?
Response The current wavelength in nanometers (nm) in the “9999.99”
format.
Example SENS:POW:WAVE?
1310.12
Remote Control (SCPI) Commands
High-Speed Power Meter 139
Specific Commands
SENSe:POWer:WAVElength
Description This command selects a new operating wavelength (for the
selected channel when using a PM-1620). If this command is
sent while an acquisition is in progress, the “Acquisition already
running” error message will be raised.
Syntax SENS:POW:WAVE<space><numeric_value> [<space>NM]
Parameters The <numeric_value> parameter is an operating wavelength
in nm. Any wavelength within the spectral range of the power
meter optical detector at 0.01 nm resolution may be selected.
See the power meter instruction manual for the exact spectral
range of each detector type. An invalid parameter will raise a
“Data type error” message.
Example SENS:POW:WAVE 1310.12
SENSe:POWer:WAVElength?
Description This query returns the currently selected wavelength (for the
selected channel when using a PM-1620).
Syntax SENS:POW:WAVE?
Response The current wavelength in nanometers (nm) in the “9999.99”
format.
Example SENS:POW:WAVE?
1310.12
Przeglądanie stron 148
1 2 ... 144 145 146 147 148 149 150 151 152 153 154 ... 161 162

Komentarze do niniejszej Instrukcji

Brak uwag

Opticon 12754 Handbücher

Bedienungsanleitungen und Benutzerhandbücher für Handheld-Mobilcomputer Opticon 12754.
Wir stellen 1 PDF-Handbücher Opticon 12754 zum kostenlosen herunterladen nach Dokumenttypen zur Verfügung Datenblatt






Weitere Produkte und Handbücher für Handheld-Mobilcomputer Opticon

Modelle Dokumententyp
H-27 Spezifikationen   Opticon H-27 PDA, 6 Seiten
H22-2D Bedienungsanleitung   Opticon H22-2D, 110 Seiten