Remote Control (SCPI) Commands
134 PM-1600
Specific Commands
SENSe:POWer:RANGe:HIGH
Description This command sets the power measurement range to manual
high (for the currently selected channel when using a PM-1620).
If this command is sent while an acquisition is in progress, the
Acquisition already running” error message will be raised.
Syntax SENS:POW:RANG:HIGH<space><numeric_value>
Parameters The <numeric_value> parameter can be 1, 2, 3, or 4
corresponding to manual HR1 to HR4.
Example SENS:POW:RANG:HIGH 2
See also SENS:POW:RANG?, SENS:POW:RANG:AUTO,
SENS:POW:RANG:AUTO?, SENS:POW:RANG:LIST?, and
SENS:POW:RANG:LOW
SENSe:POWer:RANGe:LIST?
Description This query returns the list of all available measurement ranges,
expressed in dBm or W depending on the current unit.
Syntax SENS:POW:RANG:LIST?
Response The list of available measuring ranges (in dBm or W) in the
“±99.99 to ±99.99” (dBm) or “+999.9999E±99 to
+999.9999E±99” (W) format (LR1..6, HR1..4).
Example SENS3:POW:RANG:LIST?
See also SENS:POW:RANG?, SENS:POW:RANG:AUTO,
SENS:POW:RANG:AUTO?, SENS:POW:RANG:HIGH, and
SENS:POW:RANG:LOW
Remote Control (SCPI) Commands
134 PM-1600
Specific Commands
SENSe:POWer:RANGe:HIGH
Description This command sets the power measurement range to manual
high (for the currently selected channel when using a PM-1620).
If this command is sent while an acquisition is in progress, the
Acquisition already running” error message will be raised.
Syntax SENS:POW:RANG:HIGH<space><numeric_value>
Parameters The <numeric_value> parameter can be 1, 2, 3, or 4
corresponding to manual HR1 to HR4.
Example SENS:POW:RANG:HIGH 2
See also SENS:POW:RANG?, SENS:POW:RANG:AUTO,
SENS:POW:RANG:AUTO?, SENS:POW:RANG:LIST?, and
SENS:POW:RANG:LOW
SENSe:POWer:RANGe:LIST?
Description This query returns the list of all available measurement ranges,
expressed in dBm or W depending on the current unit.
Syntax SENS:POW:RANG:LIST?
Response The list of available measuring ranges (in dBm or W) in the
“±99.99 to ±99.99” (dBm) or “+999.9999E±99 to
+999.9999E±99” (W) format (LR1..6, HR1..4).
Example SENS3:POW:RANG:LIST?
See also SENS:POW:RANG?, SENS:POW:RANG:AUTO,
SENS:POW:RANG:AUTO?, SENS:POW:RANG:HIGH, and
SENS:POW:RANG:LOW