Exfo PM-1600 High-Speed Power Meter Instrukcja Użytkownika Strona 100

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 99
Maintenance
90 PM-1600
Cleaning Connectors Equipped with EUI/EUA Adapters
Cleaning Connectors Equipped with EUI/EUA
Adapters
Regular cleaning of connectors equipped with EUI/EUA adapters will help
maintain optimum performance. There is no need to disassemble the unit.
To clean connectors,
1. Remove the EUI/EUA adapter from the module to expose the optical
connector baseplate and ferrule.
2. Use a lint-free wiping cloth and deposit only one drop of isopropyl
alcohol on it.
IMPORTANT
If any damage occurs to internal connectors, the module casing will
have to be opened and a new calibration will be required.
IMPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the wiping cloth, and do not use bottles
that distribute too much alcohol at a time.
Push
Turn
Pull
1
2
3
Maintenance
90 PM-1600
Cleaning Connectors Equipped with EUI/EUA Adapters
Cleaning Connectors Equipped with EUI/EUA
Adapters
Regular cleaning of connectors equipped with EUI/EUA adapters will help
maintain optimum performance. There is no need to disassemble the unit.
To clean connectors,
1. Remove the EUI/EUA adapter from the module to expose the optical
connector baseplate and ferrule.
2. Use a lint-free wiping cloth and deposit only one drop of isopropyl
alcohol on it.
IMPORTANT
If any damage occurs to internal connectors, the module casing will
have to be opened and a new calibration will be required.
IMPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between
the tip of the bottle and the wiping cloth, and do not use bottles
that distribute too much alcohol at a time.
Push
Turn
Pull
1
2
3
Przeglądanie stron 99
1 2 ... 95 96 97 98 99 100 101 102 103 104 105 ... 161 162

Komentarze do niniejszej Instrukcji

Brak uwag