Exfo IQS-12004B DWDM Passive Component Test System for Instrukcja Użytkownika Strona 369

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 368
COM Objects Reference
DWDM Passive Component Test System 357
Analysis Interface—Methods
Analysis
SetUserWavelengthRange
Description This method sets a user-specific wavelength range for the
analysis of the results.
Syntax object.SetUserWavelengthRange (LowerBound,UpperBound)
Parameters LowerBound: Required. A Double value indicating the starting
wavelength to limit the range for the analysis features.
UpperBound: Required. A Double value indicating the ending
wavelength to limit the range for the analysis features.
Response None
Note This method allows you to specify a second wavelength range
for the results of the analysis. An error will be raised if the
LowerBound value is higher than the UpperBound value.
The results may be obtained through the following functions:
FlatnessOnUserRange, IlOnUserRange, OrlOnUserRange and
PdlOnUserRange.
Przeglądanie stron 368
1 2 ... 364 365 366 367 368 369 370 371 372 373 374 ... 436 437

Komentarze do niniejszej Instrukcji

Brak uwag