SCPI Commands Reference
264 IQS-12004B
Specific Commands
PolarizationStateAdjuster:StateOfPolarization
Description This command sets a new polarization state on the IQS-5150
polarization state adjuster.
Syntax PSA:SOP<wsp><StateOfPolarization>
Parameter The <StateOfPolarization> parameter value is a new state of
polarization. It can be:
0, 1, 2, or 3
Response None
Example PSA:SOP 2
PolarizationStateAdjuster:StateOfPolarization?
Description The response to this query indicates the current polarization
state on the IQS-5150 polarization state adjuster.
Syntax PSA:SOP?
Parameter None
Response A string response representing the current state of polarization.
It can be:
0, 1, 2, or 3
Example PSA:SOP?