User GuideHigh Performance Power Meterand Optical Head for IQS-600IQS-1700
Introducing the IQS-1700 High Performance Power Meter2 IQS-1700Main FeaturesIt features a high sampling rate and fast stabilization, and comes in one-
Maintenance92 IQS-1700Cleaning Detector PortsCleaning Detector PortsRegular cleaning of detectors will help maintain measurement accuracy.To clean det
MaintenanceHigh Performance Power Meter 93Recalibrating the UnitRecalibrating the UnitEXFO manufacturing and service center calibrations are based on
Maintenance94 IQS-1700Recycling and Disposal (Applies to European Union Only)To help you with calibration follow-up, EXFO provides a special calibrati
High Performance Power Meter 9512 TroubleshootingSolving Common ProblemsProblem Probable Cause Recommended ActionLED push button does not illuminate.P
Troubleshooting96 IQS-1700Solving Common Problems* * * * * * * is displayed as power value. Channel not active. Channel not selected for Min. / Max.
TroubleshootingHigh Performance Power Meter 97Viewing Online DocumentationViewing Online DocumentationAn online version of the IQS-1700 High Performan
Troubleshooting98 IQS-1700Contacting the Technical Support GroupContacting the Technical Support GroupTo obtain after-sales service or technical suppo
TroubleshootingHigh Performance Power Meter 99Contacting the Technical Support GroupYou may also be requested to provide software and module version n
Troubleshooting100 IQS-1700TransportationTransportationMaintain a temperature range within specifications when transporting the unit. Transportation d
High Performance Power Meter 10113 WarrantyGeneral InformationEXFO Inc. (EXFO) warrants this equipment against defects in material and workmanship for
Introducing the IQS-1700 High Performance Power MeterHigh Performance Power Meter 3Main FeaturesThe graphical display mode shows all channels on a rea
Warranty102 IQS-1700ExclusionsExclusionsEXFO reserves the right to make changes in the design or construction of any of its products at any time witho
WarrantyHigh Performance Power Meter 103Service and RepairsService and RepairsEXFO commits to providing product service and repair for five years foll
Warranty104 IQS-1700EXFO Service Centers WorldwideEXFO Service Centers WorldwideIf your product requires servicing, contact your nearest authorized se
High Performance Power Meter 105A Technical SpecificationsIMPORTANTThe following technical specifications can change without notice. The information p
High Performance Power Meter 107B SCPI Command ReferenceThis appendix presents detailed information on the commands and queries supplied with your IQS
SCPI Command Reference108 IQS-1700Quick Reference Command TreeEXTRema <Extrema> 118EXTRema? 119MEASure[1..n] [SCALar] POWer MAXimum? 120MINimum?
SCPI Command ReferenceHigh Performance Power Meter 109Quick Reference Command Tree[MAGNitude]? [MINimum|MAXimum|DEFault] 144OFFSet [MAGNitude] <Cor
SCPI Command Reference110 IQS-1700Quick Reference Command TreeWAVelength? [MINimum|MAXimum|DEFault] 170SLINstrument CATalog? 171CATalog FULL? 172SNUMb
SCPI Command ReferenceHigh Performance Power Meter 111Product-Specific Commands—DescriptionProduct-Specific Commands—Description:ABORt[1..n]Descriptio
Introducing the IQS-1700 High Performance Power Meter4 IQS-1700Typical Appl icationsTypical ApplicationsYour power meter is suitable for numerous appl
SCPI Command Reference112 IQS-1700Product-Specific Commands—Description:FETCh[1..n][:SCALar]:POWer:DC?Description This query returns the stored value
SCPI Command ReferenceHigh Performance Power Meter 113Product-Specific Commands—Description:FORMat[1..n][:DATA]Description This command changes the re
SCPI Command Reference114 IQS-1700Product-Specific Commands—Description:FORMat[1..n][:DATA]?Description This query returns the resolution of the power
SCPI Command ReferenceHigh Performance Power Meter 115Product-Specific Commands—Description:INITiate[1..n]:AUTODescription This command starts or stop
SCPI Command Reference116 IQS-1700Product-Specific Commands—DescriptionThis parameter allows to set the acquisition mode:CONT set Continuous acquisiti
SCPI Command ReferenceHigh Performance Power Meter 117Product-Specific Commands—Description:INITiate[1..n]:AUTO?Description This query returns a value
SCPI Command Reference118 IQS-1700Product-Specific Commands—Description:INITiate[1..n]:EXTRemaDescription This command starts or stops the Min./Max. p
SCPI Command ReferenceHigh Performance Power Meter 119Product-Specific Commands—Description:INITiate[1..n]:EXTRema?Description This query returns a va
SCPI Command Reference120 IQS-1700Product-Specific Commands—Description:MEASure[1..n][:SCALar]:POWer:MAXimum?Description This query returns the maximu
SCPI Command ReferenceHigh Performance Power Meter 121Product-Specific Commands—Description:MEASure[1..n][:SCALar]:POWer:MINimum?Description This quer
Introducing the IQS-1700 High Performance Power MeterHigh Performance Power Meter 5ConventionsConventionsBefore using the product described in this gu
SCPI Command Reference122 IQS-1700Product-Specific Commands—Description:MMEMory[1..n]:ACQuisitionDescription This command initiates a data acquisition
SCPI Command ReferenceHigh Performance Power Meter 123Product-Specific Commands—DescriptionThe acquisition state can be modified with:CONT -sets the C
SCPI Command Reference124 IQS-1700Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition?Description This query returns the acquisition flag.
SCPI Command ReferenceHigh Performance Power Meter 125Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition:DURationDescription This command
SCPI Command Reference126 IQS-1700Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition:DURation?Description This query returns the duration
SCPI Command ReferenceHigh Performance Power Meter 127Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition:DURation:MMAXimumDescription Thi
SCPI Command Reference128 IQS-1700Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition:DURation:MMAXimum?Description This query returns the
SCPI Command ReferenceHigh Performance Power Meter 129Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition:DURation:MMAXimum:STATeDescripti
SCPI Command Reference130 IQS-1700Product-Specific Commands—Description:MMEMory[1..n]:ACQuisition:DURation:MMAXimum:STATe?Description This query retur
SCPI Command ReferenceHigh Performance Power Meter 131Product-Specific Commands—Description:MMEMory[1..n]:FNAMeDescription This command is used to set
SCPI Command Reference132 IQS-1700Product-Specific Commands—Description:MMEMory[1..n]:FNAMe?Description This query returns the acquisition file's
SCPI Command ReferenceHigh Performance Power Meter 133Product-Specific Commands—Description:READ[1..n][:SCALar]:POWer:DC?Description With this query,
SCPI Command Reference134 IQS-1700Product-Specific Commands—DescriptionExample(s) READ:SCAL:POW:DC? Returns -1.254000E+001READ:SCAL:POW:DC? Returns 92
SCPI Command ReferenceHigh Performance Power Meter 135Product-Specific Commands—Description:SENSe[1..n]:AVERage[:STATe]Description This command turns
SCPI Command Reference136 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:AVERage[:STATe]?Description This query returns the current averagi
SCPI Command ReferenceHigh Performance Power Meter 137Product-Specific Commands—Description:SENSe[1..n]:AVERage:COUNtDescription Sets the number of me
SCPI Command Reference138 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:AVERage:COUNt?Description This query returns the number of measure
SCPI Command ReferenceHigh Performance Power Meter 139Product-Specific Commands—DescriptionResponse(s) AverageCount: The response data syntax for <
SCPI Command Reference140 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:CORRection:COLLect:ZERODescription This command performs an offset
SCPI Command ReferenceHigh Performance Power Meter 141Product-Specific Commands—Description:SENSe[1..n]:CORRection:COLLect:ZERO:ALLDescription This co
High Performance Power Meter 72 Safety InformationYour power meter does not contain laser components in itself. However, you will be using it with lig
SCPI Command Reference142 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:CORRection:FACTor[:MAGNitude]Description This command sets a corre
SCPI Command ReferenceHigh Performance Power Meter 143Product-Specific Commands—DescriptionExample(s) SENS:CORR:FACT:MAGN 2Notes The correction factor
SCPI Command Reference144 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:CORRection:FACTor[:MAGNitude]?Description This query returns the c
SCPI Command ReferenceHigh Performance Power Meter 145Product-Specific Commands—DescriptionResponse(s) CorrectionFactor: The response data syntax for
SCPI Command Reference146 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]Description This command sets an offs
SCPI Command ReferenceHigh Performance Power Meter 147Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <
SCPI Command Reference148 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]?Description This query returns the o
SCPI Command ReferenceHigh Performance Power Meter 149Product-Specific Commands—DescriptionResponse(s) CorrectionOffset: The response data syntax for
SCPI Command Reference150 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:FREQuency:CONTinuousDescription This command sets the continuous a
SCPI Command ReferenceHigh Performance Power Meter 151Product-Specific Commands—Description:SENSe[1..n]:FREQuency:CONTinuous?Description This query re
SCPI Command Reference152 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:FREQuency:CONTinuous:CATalog?Description This query returns the li
SCPI Command ReferenceHigh Performance Power Meter 153Product-Specific Commands—Description:SENSe[1..n]:FREQuency:NCONtinuousDescription This command
SCPI Command Reference154 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:FREQuency:NCONtinuous?Description This query returns the current s
SCPI Command ReferenceHigh Performance Power Meter 155Product-Specific Commands—Description:SENSe[1..n]:FREQuency:NCONtinuous:CATalog?Description This
SCPI Command Reference156 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:RANGe:AUTODescription This command enables or disables
SCPI Command ReferenceHigh Performance Power Meter 157Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:RANGe:AUTO?Description This query r
SCPI Command Reference158 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:RANGe:SCALeDescription This command sets the measuremen
SCPI Command ReferenceHigh Performance Power Meter 159Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:RANGe:SCALe?Description This query
SCPI Command Reference160 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:RANGe:SCALe:LIST?Description This query returns the lis
SCPI Command ReferenceHigh Performance Power Meter 161Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:REFerenceDescription This command s
High Performance Power Meter 93 Getting Started with Your High Performance Power MeterThis chapter contains information on how to insert and remove te
SCPI Command Reference162 IQS-1700Product-Specific Commands—DescriptionDEFault allows the instrument to select a value for the <Reference> param
SCPI Command ReferenceHigh Performance Power Meter 163Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:REFerence?Description This query re
SCPI Command Reference164 IQS-1700Product-Specific Commands—DescriptionResponse(s) Reference: The response data syntax for <Reference> is define
SCPI Command ReferenceHigh Performance Power Meter 165Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:REFerence:ALLDescription This comma
SCPI Command Reference166 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:REFerence:DISPlayDescription This command performs a ne
SCPI Command ReferenceHigh Performance Power Meter 167Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:REFerence:STATeDescription This com
SCPI Command Reference168 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:POWer[:DC]:REFerence:STATe?Description This query returns a value
SCPI Command ReferenceHigh Performance Power Meter 169Product-Specific Commands—Description:SENSe[1..n]:POWer:WAVelengthDescription This command selec
SCPI Command Reference170 IQS-1700Product-Specific Commands—Description:SENSe[1..n]:POWer:WAVelength?Description This query returns the currently sele
SCPI Command ReferenceHigh Performance Power Meter 171Product-Specific Commands—Description:SLINstrument:CATalog?Description This query returns a comm
Getting Started with Your High Performance Power Meter10 IQS-1700Inserting and Removing Test ModulesTo insert a module into the controller or expansio
SCPI Command Reference172 IQS-1700Product-Specific Commands—Description:SLINstrument:CATalog:FULL?Description This query returns a list of <STRING
SCPI Command ReferenceHigh Performance Power Meter 173Product-Specific Commands—Description:SNUMber?Description This query returns a value indicating
SCPI Command Reference174 IQS-1700Product-Specific Commands—Description:STATus?Description This query returns a value indicating the status of the mod
SCPI Command ReferenceHigh Performance Power Meter 175Product-Specific Commands—Description:STATus:OPERation:BIT[1..n]:CONDition?Description This quer
SCPI Command Reference176 IQS-1700Product-Specific Commands—Description:TRACe[1..n][:DATA]?Description Returns all points in a trace.Syntax :TRACe[1..
SCPI Command ReferenceHigh Performance Power Meter 177Product-Specific Commands—DescriptionEach power value represents a point in the trace and is ret
SCPI Command Reference178 IQS-1700Product-Specific Commands—Description:TRACe[1..n]:MAX?Description This query returns the maximum value for the speci
SCPI Command ReferenceHigh Performance Power Meter 179Product-Specific Commands—DescriptionExample(s) INIT:EXTR 1INIT:EXTR 0TRAC:MAX? TRC1Notes A trac
SCPI Command Reference180 IQS-1700Product-Specific Commands—Description:TRACe[1..n]:MIN?Description This query returns minimum value in the trace.Synt
SCPI Command ReferenceHigh Performance Power Meter 181Product-Specific Commands—DescriptionExample(s) INIT:EXTR 1INIT:EXTR 0TRAC:MIN? TRC1Notes A trac
Getting Started with Your High Performance Power MeterHigh Performance Power Meter 11Inserting and Removing Test Modules4. Insert the protruding edges
SCPI Command Reference182 IQS-1700Product-Specific Commands—Description:TRACe[1..n]:POINtsDescription This command sets the number of points to store
SCPI Command ReferenceHigh Performance Power Meter 183Product-Specific Commands—DescriptionThis parameter is used to select the trace to use. Each tra
SCPI Command Reference184 IQS-1700Product-Specific Commands—Description:TRACe[1..n]:POINts?Description This query returns the number of points in a sp
SCPI Command ReferenceHigh Performance Power Meter 185Product-Specific Commands—DescriptionExample(s) TRAC:POIN? TRC1Notes The suffix of TRACe[1..n ]
SCPI Command Reference186 IQS-1700Product-Specific Commands—Description:UNIT[1..n]:POWerDescription The parameter corresponds to the power units for t
SCPI Command ReferenceHigh Performance Power Meter 187Product-Specific Commands—Description:UNIT[1..n]:POWer?Description This query returns the curren
IndexHigh Performance Power Meter 189IndexAAbout function tab ... 99absolute power measurements ...
Index190 IQS-1700displayresolution ... 28type ... 76div
IndexHigh Performance Power Meter 191measuringabsolute power... 43corrected power ...
ii IQS-1700Copyright © 2002–2014 EXFO Inc. All rights reserved. No part of this publication may be reproduced, stored in a retrieval system or transmi
Getting Started with Your High Performance Power Meter12 IQS-1700Inserting and Removing Test ModulesThe module is correctly inserted when its front pa
Index192 IQS-1700Ssafetycaution ... 5conventions ... 5warni
IndexHigh Performance Power Meter 193WW/W correction factor... 48warrantycertification ...
NOTICE抩⛙CHINESE REGULATION ON RESTRICTION OF HAZARDOUS SUBSTANCES₼⦌␂ℝ☀⹂䓸德棟Ⓟ䤓屓⸩NAMES AND CONTENTS OF THE TOXIC OR HAZARDOUS SUBSTANCES OR ELEMENTS CONT
MARKING REQUIREMENTS㪖㽷尐㻑Productℶ❐Environmental protection use period (years)䘾⬒≬㔳∎䞷㦮棟 ( )Logo㪖㉦This E;)2 product㦻 EXFO ℶ❐10Batterya䟄㻯aa. If applicabl
www.EXFO.com · [email protected]/N: 1066977CORPORATE HEADQUARTERS 400 Godin Avenue Quebec (Quebec) G1M 2K2 CANADATel.: 1 418 683-0211 · Fax: 1 418 683-
Getting Started with Your High Performance Power MeterHigh Performance Power Meter 13Inserting and Removing Test Modules3. Cover empty slots with the
Getting Started with Your High Performance Power Meter14 IQS-1700Connecting the Optical Head to the Power MeterConnecting the Optical Head to the Powe
Getting Started with Your High Performance Power MeterHigh Performance Power Meter 15Securing the Optical Head onto Your Work SurfaceSecuring the Opti
Getting Started with Your High Performance Power Meter16 IQS-1700Starting the High Performance Power Meter ApplicationTo start the application:1. From
Getting Started with Your High Performance Power MeterHigh Performance Power Meter 17Starting the High Performance Power Meter ApplicationThe main win
Getting Started with Your High Performance Power Meter18 IQS-1700Starting the High Performance Power Meter ApplicationTitle BarThe title bar is locate
Getting Started with Your High Performance Power MeterHigh Performance Power Meter 19Entering Values Using Sliders and Numeric BoxesEntering Values Us
Getting Started with Your High Performance Power Meter20 IQS-1700Exiting the ApplicationExiting the ApplicationClosing any application that is not cur
High Performance Power Meter 214 Setting Up Your High Performance Power MeterYou can set the following parameters on your IQS-1700: Channel display
ContentsHigh Performance Power Meter iiiContentsCertification Information ...
Setting Up Your High Performance Power Meter22 IQS-1700Setting Channel DisplaySetting Channel Display The channel display allows you to select which c
Setting Up Your High Performance Power MeterHigh Performance Power Meter 23Selecting a Channel (Multi-Channel Models)Selecting a Channel (Multi-Channe
Setting Up Your High Performance Power Meter24 IQS-1700Naming ChannelsNaming ChannelsA user-selected name can be given to each power meter channel. Th
Setting Up Your High Performance Power MeterHigh Performance Power Meter 25Selecting the WavelengthSelecting the WavelengthWhen taking accurate measur
Setting Up Your High Performance Power Meter26 IQS-1700Managing Wavelength ListsManaging Wavelength ListsThe wavelengths you want to use with your IQS
Setting Up Your High Performance Power MeterHigh Performance Power Meter 27Selecting the Measurement UnitTo delete a wavelength from the list:1. From
Setting Up Your High Performance Power Meter28 IQS-1700Setting the Display ResolutionSetting the Display ResolutionDepending on the required resolutio
Setting Up Your High Performance Power MeterHigh Performance Power Meter 29Setting the Display ResolutionIf you have previously selected watts as unit
Setting Up Your High Performance Power Meter30 IQS-1700Setting the Refresh RateSetting the Refresh RateThis function allows you to define the refresh
Setting Up Your High Performance Power MeterHigh Performance Power Meter 31Setting the Measurement RangeSetting the Measurement RangeThe measurement r
Contentsiv IQS-17007 Recording Power Signal Variations ...538 Performing Acquisitions ...
Setting Up Your High Performance Power Meter32 IQS-1700Setting the Measurement RangeTo set the measurement range:1. Click the Settings function tab, t
Setting Up Your High Performance Power MeterHigh Performance Power Meter 33Setting the Measurement Range4. Use the arrow buttons next to the Scale lis
Setting Up Your High Performance Power Meter34 IQS-1700Saving and Recalling ConfigurationsSaving and Recalling ConfigurationsOnce you have set the IQS
Setting Up Your High Performance Power MeterHigh Performance Power Meter 35Saving and Recalling ConfigurationsTo recall a configuration:1. Select the
High Performance Power Meter 375 Preparing Your High Performance Power Meter for a TestThe OHS-1700 Optical Head can be used with connectorized or non
Preparing Your High Performance Power Meter for a Test38 IQS-1700Cleaning and Connecting Optical Fibers3. Carefully align the connector and port to pr
Preparing Your High Performance Power Meter for a TestHigh Performance Power Meter 39Nulling OffsetsNulling OffsetsTemperature and humidity variations
Preparing Your High Performance Power Meter for a Test40 IQS-1700Nulling OffsetsTo perform an offset nulling on one channel:1. Install the protective
Preparing Your High Performance Power Meter for a TestHigh Performance Power Meter 41Nulling OffsetsTo perform an offset nulling on all channels (two-
ContentsHigh Performance Power Meter vIndex ...
Preparing Your High Performance Power Meter for a Test42 IQS-1700Nulling Offsets3. Click OK to perform the offset nulling, or Cancel to exit. A red me
High Performance Power Meter 436 Measuring PowerPower measurements can be displayed in two ways: absolute relativeIt is also possible to use a corre
Measuring Power44 IQS-1700Displaying Absolute PowerTo display absolute power:1. Select the Instrument function tab or the Channels tab of the Settings
Measuring PowerHigh Performance Power Meter 45Measuring Relative PowerMeasuring Relative PowerPower measurements can be displayed as a deviation from
Measuring Power46 IQS-1700Selecting the Reference ValueSelecting the Reference ValueThe reference value influences your measurements once selected and
Measuring PowerHigh Performance Power Meter 47Editing the Reference ListEditing the Reference ListThe Reference list can be changed to adapt to your t
Measuring Power48 IQS-1700Measuring Corrected PowerTo delete a user reference from the list:1. From the Settings function tab, select the Lists tab.2.
Measuring PowerHigh Performance Power Meter 49Measuring Corrected PowerTo set a correction factor:1. From the Settings function tab, select the Channe
Measuring Power50 IQS-1700Using the Offset FunctionUsing the Offset FunctionThe offset function is used when you want to take into account, in the pow
Measuring PowerHigh Performance Power Meter 51Averaging MeasurementsAveraging MeasurementsWhen the averaging function is enabled on the IQS-1700 High
Certification Informationvi IQS-1700Certification InformationNorth America Regulatory StatementThis unit was certified by an agency approved in both C
Measuring Power52 IQS-1700Averaging MeasurementsNote: When measurement conditions change, the number of the acquisition is reset to zero in order to s
High Performance Power Meter 537 Recording Power Signal VariationsThe Min./Max. function allows you to record the extremes of a varying power signal w
Recording Power Signal Variations54 IQS-17002. Select the channel for which you want to see the power variation data on-screen by clicking the corresp
Recording Power Signal VariationsHigh Performance Power Meter 55The results of the Min./Max. function are continuously updated in the matching lists.
High Performance Power Meter 578 Performing AcquisitionsYou can perform your acquisition on one or several channels at the same time (in the case of m
Performing Acquisitions58 IQS-1700Selecting the Sampling TypeSelecting the Sampling TypeYou can perform different types of samplings with your power m
Performing AcquisitionsHigh Performance Power Meter 59Selecting the Sampling TypeTo select a sampling type:1. From the Settings function tab, select t
Performing Acquisitions60 IQS-1700Selecting the Sampling TypeThe acquisition rate applies to all channels when using a multichannel power meter.You ca
Performing AcquisitionsHigh Performance Power Meter 61Editing the Frequency Divider ListEditing the Frequency Divider ListTo select the frequency at w
Certification InformationHigh Performance Power Meter viiEuropean Community Declaration of ConformityAn electronic version of the declaration of confo
Performing Acquisitions62 IQS-1700Selecting the Acquisition ModeSelecting the Acquisition ModeYou can use three data acquisition modes, regardless of
Performing AcquisitionsHigh Performance Power Meter 63Selecting the Acquisition ModeTo select the acquisition mode:1. From the Settings function tab,
Performing Acquisitions64 IQS-1700Selecting the Acquisition ModeSetting Up Timed Acquisition DurationA timed acquisition starts when you start the pro
Performing AcquisitionsHigh Performance Power Meter 65Selecting the Acquisition ModeSetting Up Delayed Acquisition DelayDelayed acquisition starts at
Performing Acquisitions66 IQS-1700Selecting the Acquisition ModeSetting Trigger Acquisition ParametersConditional data acquisition can be performed us
Performing AcquisitionsHigh Performance Power Meter 67Selecting the Acquisition ModeTo set up the power level trigger condition:1. From Trigger of the
Performing Acquisitions68 IQS-1700Starting the AcquisitionStarting the AcquisitionOnce you have set your parameters, you can start the acquisition. To
Performing AcquisitionsHigh Performance Power Meter 69Starting the AcquisitionYou can also start the acquisition in the Graph tab by clicking the Star
Performing Acquisitions70 IQS-1700Consulting Acquired DataConsulting Acquired DataOnce you have acquired data, it is possible to view the results in I
Performing AcquisitionsHigh Performance Power Meter 71Consulting Acquired Data3. In the viewer, retrieve the corresponding file using the button. T
Performing Acquisitions72 IQS-1700Consulting Acquired Data To view your acquisition data, select the Data tab.
Performing AcquisitionsHigh Performance Power Meter 73Consulting Acquired Data To view the graph corresponding to your acquisition data, select the G
Performing Acquisitions74 IQS-1700Consulting Acquired DataIf you want to save your acquisition file as a text file, use the Export button located on t
High Performance Power Meter 759 Performing and Analyzing Graph AcquisitionsThe Graph mode of your power meter allows you to view your acquisition as
Performing and Analyzing Graph Acquisitions76 IQS-1700Setting Up Graph ParametersNote: You can only hide or display channels actually used for this ac
Performing and Analyzing Graph AcquisitionsHigh Performance Power Meter 77Setting Up Graph ParametersTo change the base unit of the acquisition (dBm o
Performing and Analyzing Graph Acquisitions78 IQS-1700Printing Graph ResultsPrinting Graph ResultsOnce you have acquired data and displayed a graph, i
Performing and Analyzing Graph AcquisitionsHigh Performance Power Meter 79Using the Zoom FunctionUsing the Zoom FunctionOnce you have performed an acq
Performing and Analyzing Graph Acquisitions80 IQS-1700Using the Zoom Function allows you to enlarge the trace display at the precise location where
Performing and Analyzing Graph AcquisitionsHigh Performance Power Meter 81Displaying and Moving MarkersDisplaying and Moving MarkersOnce you have acqu
High Performance Power Meter 11 Introducing the IQS-1700 High Performance Power MeterDesigned for the IQS-600 Integrated Qualification System, the IQS
Performing and Analyzing Graph Acquisitions82 IQS-1700Displaying and Moving MarkersNote: If you have enabled the markers and you move to another tab (
High Performance Power Meter 8310 Monitoring Power Meter ModulesWhen using your IQS-1700 High Performance Power Meter module, either alone or with oth
Monitoring Power Meter Modules84 IQS-1700Using Monitor WindowsTo select modules and display their monitor windows:1. On the Current Modules function t
Monitoring Power Meter ModulesHigh Performance Power Meter 85Using Monitor Windows3. Click Start Monitor to apply your selection.IQS Manager will disp
Monitoring Power Meter Modules86 IQS-1700Using QuickToolsUsing QuickToolsWith QuickTools, you can fine-tune your module directly, while keeping an eye
Monitoring Power Meter ModulesHigh Performance Power Meter 87Using QuickToolsTo cl ose Q uick To o l s:Click the Close button located at the top of th
High Performance Power Meter 8911 MaintenanceTo help ensure long, trouble-free operation: Always inspect fiber-optic connectors before using them and
Maintenance90 IQS-1700Cleaning Fixed ConnectorsCleaning Fixed ConnectorsRegular cleaning of connectors will help maintain optimum performance. Do not
MaintenanceHigh Performance Power Meter 91Cleaning Fixed Connectors5. Throw out the wiping cloths after one use.6. Moisten a cleaning tip (2.5 mm tip)
Komentarze do niniejszej Instrukcji