TECHNICAL SPECIFICATIONS
Model
a
FTB-7400E-XXXX FTB-7400E-CWO FTB-7400E-CWE FTB-7400E-CWS FTB-7400E-CWCL
Wavelengths (nm)
b
1310 ± 20
1383 ± 1
1550 ± 20
1625 ± 10
1270 ± 3
1290 ± 3
1310 ± 3
1330 ± 3
1350 ± 3
1410 ± 3
1430 ± 3
1450 ± 3
1470 ± 3
1490 ± 3
1510 ± 3
1530 ± 3
1550 ± 3
1570 ± 3
1590 ± 3
1610 ± 3
Dynamic range at 20 +s (dB)
c
42/40/41/41 41/41/41/41 41/41/41/41 41/41/ 41/41 41/41/ 40/40
Event dead zone (m)
d
0.8 0.8 0.8 0.8 0.8
Attenuation dead zone (m)
d
4/4/4.5/4.5 4.5 4.5 4.5 4.5
Distance range (km) 1.25, 2.5, 5, 10, 20, 40,
80, 160, 260, 400
1.25, 2.5, 5, 10, 20, 40,
80, 160, 260, 400
1.25, 2.5, 5, 10, 20, 40,
80, 160, 260, 400
1.25, 2.5, 5, 10, 20, 40,
80, 160, 260, 400
1.25, 2.5, 5, 10, 20, 40,
80, 160, 260, 400
Pulse width (ns) 5, 10, 30, 100, 275, 1000,
2500, 10 000, 20 000
5, 10, 30, 100, 275, 1000,
2500, 10 000, 20 000
5, 10, 30, 100, 275, 1000,
2500, 10 000, 20 000
5, 10, 30, 100, 275, 1000,
2500, 10 000, 20 000
5, 10, 30, 100, 275, 1000,
2500, 10 000, 20 000
Linearity (dB/dB)
b
±0.03 ±0.03 ±0.03 ±0.03 ±0.03
Loss threshold (dB) 0.01 0.01 0.01 0.01 0.01
Loss resolution (dB) 0.001 0.001 0.001 0.001 0.001
Sampling resolution (m) 0.04 to 5 0.04 to 5 0.04 to 5 0.04 to 5 0.04 to 5
Sampling points Up to 256 000 Up to 256 000 Up to 256 000 Up to 256 000 Up to 256 000
Distance uncertainty (m)
e
±(0.75 + 0.001 % x
distance + sampling
resolution)
±(0.75 + 0.001 % x
distance + resolution)
±(0.75 + 0.001 % x
distance + resolution)
±(0.75 + 0.001 % x
distance + sampling
resolution)
±(0.75 + 0.001 % x
distance + sampling
resolution)
Measurement time
User-defined
(5 sec. minimum to
60 min. maximum)
User-defined
(60 min. maximum)
User-defined
(60 min. maximum)
User-defined
(5 sec. minimum to
60 min. maximum)
User-defined
(5 sec. minimum to
60 min. maximum)
Typical real-time refresh (Hz) 44444
Stable source output
power (dBm)
f
–4.5 (7400E-0023B) –4.5 –4.5 –4.5 –4.5
Komentarze do niniejszej Instrukcji