Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Instrukcja Użytkownika Strona 587

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 586
SCPI Command Reference
OSA 581
Product-Specific Commands—Description
The <Data> response corresponds to the Y-axis
values vector of the trace. Returned values are
evenly spaced relative to the X-axis expressed in
meter.
X-axis wavelength interval between each Y value
is determined as follow:
interval = (stop - start) / (count - 1) where:
start = TRACe[:DATA]:X:STARt[:WAVelength]?
stop = TRACe[:DATA]:X:STOP[:WAVelength]?
count = TRACe:POINts?
The points unit is determined by the trace
definition context. When trace data represents
absolute power, returned values are in dBm.
When trace data represents relative power,
returned values are in dB.
Example(s) <Do measurement>
FORMat:DATA ASC
TRAC:DATA? "TRC1" Returns
-5.246202E+001,-5.246195E+001,-5.246181E+001
,...
FORMat:DATA PACK
TRAC:DATA? "TRC1" Returns binary data
:TRACe[:DATA][:Y][:WAVElength]?
Przeglądanie stron 586
1 2 ... 582 583 584 585 586 587 588 589 590 591 592 ... 637 638

Komentarze do niniejszej Instrukcji

Brak uwag