Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Instrukcja Użytkownika Strona 455

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 454
SCPI Command Reference
OSA 449
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
SELect
Description This command sets the name of the selected
WDM channel analysis setup.
At *RST, there is no selection: a single null string
is returned.
Syntax :CALCulate[1..n][:WDM]:CHANnel:SELect<wsp
><Select>
Parameter(s) Select:
The program data syntax for <Select> is defined
as a <STRING PROGRAM DATA> element.
The <Select> parameter corresponds to the
name of the channel setup to select. The channel
name cannot be empty.
Example(s) CALC:WDM:CHAN:DEF "C_001",1525.000 NM
CALC:WDM:CHAN:SEL "C_001"
CALC:WDM:CHAN:SEL? Returns "C_001"
See Also :CALCulate[1..n][:WDM]:CHANnel:CATalog?
:CALCulate[1..n][:WDM]:CHANnel[:DEFine]
:CALCulate[1..n][:WDM]:CHANnel:SELect?
:CALCulate[1..n][:WDM]:CHANnel:NSELect