SCPI Command Reference
494 FTB-5240S/S-P/BP
Product-Specific Commands—Description
Response(s) Signal:
The response data syntax for <Signal> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Signal> response corresponds to the
computed signal power level.
Example(s) CALC:CHAN:DEL:ALL
CALC:CHAN:DEF "ITU_019", 229.7860 THZ
CALC:CHAN:SEL "ITU_019"
CALC:CHAN:SIGP:TYPE TPOW
<Do measurement>
CALC:DATA:CHAN:SEL "ITU_019"
UNIT:POW DBM
CALC:DATA:CHAN:SIGP? Returns
-3.430000E+000
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO:SIGnal
Power :TYPE
:CALCulate[1..n][:WDM]:CHANnel:SIGnalPower:
TYPE
:CALCulate[1..n][:WDM]:DATA:CHANnel:SIGnalP
ower:TYPE?
:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel::STATus
:QUEStionable:BIT<9|10|11>:CONDition?
:CALCulate[1..n][:WDM]:DATA:CHANnel:
SIGnalPower?