Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 80

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 79
Setting Up the Instrument in WDM Mode
72 FTB-5240S/5240BP
Setting Up WDM Analysis Parameters
2. Select the Default Thresholds tab.
3. Enter values in the boxes as explained below:
³ Wavelength/Frequency (nm/GHz): indicates the channel’s central
wavelength/frequency.
³ Signal power (dBm): indicates the signal power for the selected
channel (excludes noise).
³ OSNR (dB): is the Optical Signal to Noise Ratio, given by Signal
power (according to the current calculation method, in dBm)
minus Noise (according to the current calculation method, in
dBm).
³ Noise (dBm): indicates the level of the noise for the selected
channel.
Przeglądanie stron 79
1 2 ... 75 76 77 78 79 80 81 82 83 84 85 ... 536 537

Komentarze do niniejszej Instrukcji

Brak uwag