SCPI Command Reference
474 FTB-5240S/5240BP
Product-Specific Commands—Description
:CALCulate[1..n]:WAVelength:OFFSet
[:OUTPut]
Description This command sets the wavelength offset value.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 0.000 nm.
Syntax :CALCulate[1..n]:WAVelength:OFFSet[:OUTPut]
<wsp><WavelegthOffset>
Parameter(s) WavelegthOffset:
The program data syntax for <WavelegthOffset>
is defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Selects the wavelength Offset value.
Range: [-99.999 … 99.999] nm
Example(s) CALC:WAV:OFFSet:OUTPut 0.000
See Also CALCulate[1..n]:WAVelength:OFFSet:OUTPut?