Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Instrukcja Użytkownika Strona 134

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 133
Setting Up the Instrument in Drift Mode
126 FTB-5240S/5240BP
Setting Up Drift Analysis Parameters
3. Under Default channel settings, define the following parameters as
needed:
³ Clear the Activate default channel selection, to use the currently
defined channel for analysis. This reduces the analysis time by
eliminating the peak detection over the complete spectral range.
The peaks outside the defined channel will not be detected or
analyzed.
³ Channel width (GHz or nm): indicates the limit inside which the
power values will be considered in the channel.
For default channels, the channel width that sets the limits of the
channel, should be the same as the channel distance or smaller
(channel distance is defined while creating a channel list). If the
channel width is not compatible with the channel spacing, either a
single peak may be found for two distinct channels and two
analysis would be performed and displayed for that peak, or, it is
possible that two peaks may be found within the same channel
Przeglądanie stron 133
1 2 ... 129 130 131 132 133 134 135 136 137 138 139 ... 536 537

Komentarze do niniejszej Instrukcji

Brak uwag