Exfo FVA-3150 Variable Attenuator Instrukcja Użytkownika Strona 95

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 94
Remote Control Commands
Variable Attenuator 89
Specific commands-Quick Reference Command Tree
Parameters The <value> parameter can be
“MIN” (to set the FVA-3150 to the minimum setting for the
current wavelength)
“MAX” (to set the FVA-3150 to the maximum setting for the
current wavelength)
a valid attenuation value followed by the units [DB]. Note
that, depending on the current measurement mode, the
<value> parameter takes different values.
In Absolute mode:
In Reference mode:
In Offset mode:
Therefore, when in Reference or Offset mode, the range of
the accepted values for the <value> parameter will change
according to the current reference or offset value.
Example INP:ATT -15.350 DB
Note This command cannot be issued while a program is running.
The ATT command is equivalent to the INP:ATT command.
Both give the same result.
value actual attenuation=
value
actual
attenuation
reference value(INP:OFFS?)=
value actual attenuation wavelength-specific offset+=
Przeglądanie stron 94
1 2 ... 90 91 92 93 94 95 96 97 98 99 100 ... 114 115

Komentarze do niniejszej Instrukcji

Brak uwag