SCPI-Befehlsreferenz
396 Serie FTB-7000
Product-Specific Commands—Description
:CONFigure[1..n]:ANAlysis:
IORefraction
Description This command sets the index of refraction that
will be used for the next acquisition.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ANAlysis:IORefraction<wsp>
<IOR>|MAXimum|MINimum|DEFault
Parameter(s) IOR:
The program data syntax for <IOR> is defined as
a <numeric_value> element. The <IOR>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
lowest supported value.
MAXimum allows to set the instrument to the
highest supported value.
DEFault allows the instrument to select a value
for the <IOR> parameter.
Sets the index of refraction.
Example(s) CONF:ANA:IOR? Ex.: Returns 1.4677
CONF:ANA:IOR 1.5
CONF:ANA:IOR? Returns 1.5