SCPI-Befehlsreferenz
372 Serie FTB-7000
Product-Specific Commands—Description
:CALCulate[1..n]:THReshold:SLOSs?
Description This query returns the splice loss threshold used
for the specified trace index.
*RST clears this value.
Syntax :CALCulate[1..n]:THReshold:SLOSs?<wsp>TRC
1|TRC2|TRC3|TRC4
Parameter(s) Label:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
TRC1|TRC2|TRC3|TRC4.
Trace index of the available wavelengths.
Response Syntax <Splice Loss>
Response(s) Splice Loss:
The response data syntax for <Splice Loss> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
Returns the splice loss threshold used by the
trace corresponding to the specified trace index.
Example(s) CONF:ANA:THR:SLOS 0.03
CONF:ACQ:MODE ACQ
INIT
INIT:STAT? Returns 0 when acquisition is
complete.
CALC:THR:SLOS? TRC1 Returns 0.03
See Also MMEMory[1..n]:LOAD:TRACe