Exfo FTB-500 (Window XP) Instrukcja Użytkownika Strona 122

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 121
Using the Optional Built-In Power Meter and VFL
112 FTB-500
Setting Thresholds and Correction Factors
You may apply a correction factor (CF) to measured power to compensate
for inaccuracies or drifts. You should change the CF after performing an
offset nulling.
where:
Power
corrected
= the corrected power value
Power
measured
= the measured power value
CF = the correction factor
For each favorite wavelength, the CF is set to 1.00 at the factory (even if
the unit indicates “----”), but allowed values range between 0.85 and
1.15.
Note: Some other products express the CF in dB, so the CF would be added to
measured power.
IMPORTANT
If you revert to factory settings, you will erase all your thresholds
and correction factors. The pass/fail status feature will be disabled
automatically.
Power
corrected
Power
measured
CF=
Przeglądanie stron 121
1 2 ... 117 118 119 120 121 122 123 124 125 126 127 ... 404 405

Komentarze do niniejszej Instrukcji

Brak uwag