Remote Control
120 FLS-2600B
Programming Commands
SOURce:SWEep:CONTinuous:PROGram?
Description This function is used to set and/or view the parameters for a
continuous sweep program, during which the application
sweeps between the minimum (start) and maximum (stop)
wavelengths.
If you only wish to view the parameters without programming
them, simply enter the query without any data following it.
Syntax SOUR:SWE:CONT:PROG?<wsp>[<start>],[<stop>],
[<speed>],[<coun>],[<rev>],[<trig>]
SOUR:SWE:CONT:PROG?
Response The <start> and <stop> parameters are the sweep start and
stop wavelengths in 9999.999 nm or 999.9999 THz format.
The <speed> parameter is the sweep speed in 99.9999 format.
The <coun> parameter is the number of passes made. The
format must be an integer from 1 to 100 or 0 (continuous).
The <rev> parameter indicates whether the reverse option is
on or off. The format must be 1|0.
The <trig> parameter indicates whether the trigger option is on
or off. The format must be 1|0.