SCPI Commands Reference
Optical Modulation Analyzer 221
Product-Specific Commands—Description
:SENSe:DDEMod:SIGNal:FREQuency
Description .This command sets the signal frequency or
wavelength. The signal spectrum needs to be
within the receiver spectrum, otherwise the
constellation chart might be distorted and noisy,
or even not visible at all. The range is from 191.5
to 196.25.
At *RST this value is set to 193.1.
Syntax
Parameter(s) Frequency:
Example(s) LINS1:SENSe:DDEMod:SIGNal:FREQuency 192
See Also SENSe:DDEMod:SIGNal:OFFSet 10