SCPI Commands Reference
200 PSO-200
Product-Specific Commands—Description
:SENSe:DDEMod:WINDow:AMPLitude
Description This command sets the measurement value for
amplitude thresholds.
At *RST value, it is set to 20%-80%.
Syntax :SENSe:DDEMod:WINDow:AMPLitude<wsp>10-
90|20-80
Parameter(s) value:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
10-90|20-80.
Sets the measurement value for amplitude
thresholds.
20-80,selects 20%-80% as the measurement
value for amplitude thresholds.
10-90 ,selects 10%-90% as the measurement
value for amplitude thresholds.
Example(s) LINS1:SENS:DDEM:WIND:AMPL 20-80
See Also SENS:DDEM:WIND:LOW 1