Przeglądanie stron 259
Especificaciones técnicas
252 MAX-610
COPPER SPECIFICATIONS
a, b, c
Transmitter characteristics
Frequency range (200 Hz to 20 kHz) Frequency resolution 1-Hz steps
Frequency uncertainty (accuracy) ±(50 ppm + 1 Hz)
Level range (dBm) –20 to 10 at 600 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance (1) 600
Frequency range (20 kHz to 2.2 MHz) Frequency resolution 1-kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range (dBm) –20 to 10 at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance (1) 100, 120, 135, 150
Frequency range (2.2 MHz to 30 MHz) Frequency resolution 1-kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range (dBm) –20 to 0 at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance (1) 100, 120, 135, 150
Receiver characteristics Reception frequency range 200 Hz to 20 kHz
20 kHz to 30 MHz
Frequency uncertainty range (accuracy) ±(50 ppm + 1 digit)
VF reception level range (dBm) –90 to 15 at 600 1
VF level uncertainty (accuracy) 200 Hz to 20 kHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
WB reception level range (dBm) –90 to 15 at 100 1 and 135 1
WB level uncertainty (accuracy)
20 kHz to 2.2 MHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
2.2 MHz to 30 MHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
Impedance (1) 100, 120, 135, 150, 600
POTS dialer DTMF 0 – 9, #, *
Phonebook 25 entries
Digital multimeter (DMM) Test type Snapshot and continuous
Impedance selection (for voltage measurement) 100 k1, 1 M1
Measurement Range Resolution Uncertainty (accuracy)
DC voltage 0 to 400 V 0.1 V for 0 to 99.9 V
1 V for 100 V to 400 V
±(|1 %| + 0.5 VDC)
AC voltage 0 to 280 Vrms 0.1 V for 0 to 99.9 V
1 V for 100 V to 280 V
±(1 % + 0.5 VAC)
Isolation resistance
(stress/leakage)
0 to 1 G1, auto-ranging
1 k1 to 99 M1
100 M1 to 999 M1
Three digits
±(2 % + 1 digit)
±(5 % + 1 digit)
Resistance 0 to 100 M1
0 to 999 1
1 k1 to 100 M1
Three digits
±(1 % + 5 1)
±(2 % + 1 digit)
Capacitance 0.1 nF to 2 μF Four digits ±(2 % + 50 pF)
DC current 0 to 110 mA 0.1 mA ±(|2 %| + 1 mA)
AC current 0 to 110 mA 0.1 mA ±(|2 %|+ 1 mA)
d
Station Ground 0 to 1 M1
0 to 999 1
1 k1 to 1 M1
Up to three digits
±(1 % + 3 1)
±(2 % + 1 digit)
NOTES
a. Subject to change without notice.
b. Typical, at 23 °C ± 3 °C, on batteries, with no type B USB connection.
c. Specifications based on 24 AWG (PE 0.5 mm) cabling.
d. From 10 mA to 110 mA.