Test Setup - Test Configurator, Timer, and System
106 FTB-700G/800 Series
BERT
Multi-Pattern Configuration
Note: Multi-Pattern Configuration is only available for DSN/PDH BERT test
application with DS1 signal (DS0 disabled) and when the pattern is
Multi-Pattern.
This feature allows sending each pattern for a specific duration,
sequentially and continuously.
Pattern represents a sequence of patterns that will be generated: 1111,
1in8, 2in8, 3in24, and QRSS.
Enable allows enabling the generation of each pattern individually in
the pattern sequence. All patterns are enabled by default. All 1’s (1111)
is disabled for Unframed test.
Individual Pattern Duration specifies the transmission duration of
each pattern: 15 s, 30 s, 45 s, 1min, 2min, 3min (default),... up to
15 min.
Restore < Test Application > Defaults
Reverts the current test application to its default factory settings.