SCPI Command Reference
340 OTDR
Product-Specific Commands—Description
:CALCulate[1..n]:ORL?
Description This query returns the value of the Optical Return
Loss measured between two markers, for the
trace corresponding to the specified trace index.
*RST clears this value.
Syntax :CALCulate[1..n]:ORL?<wsp>TRC1|TRC2|TRC3
|TRC4,<MarkerA>,<MarkerB>
Parameter(s)
³ Label:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
TRC1|TRC2|TRC3|TRC4.
Trace index of the available wavelengths.
³ MarkerA:
The program data syntax for <MarkerA> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Specifies the marker A position, in meters.
³ MarkerB:
The program data syntax for <MarkerB> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Specifies the marker B position, in meters.
Response Syntax <ORL>