Exfo FTB-7000 OTDR for FTB-200 v2 Instrukcja Użytkownika Strona 68

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 67
Testing Fibers in Advanced Mode
54 FTB-7000 Series
Setting Pass/Fail Thresholds
The following table provides the default, minimum and maximum
thresholds.
Once the thresholds are set, the application will be able to perform
Pass/Fail tests to determine the status of the various events (pass or fail).
The Pass/Fail test is performed on two occasions:
³ when analyzing or reanalyzing a trace
³ when you open a trace file
Values that are greater than the predefined thresholds are displayed in
white on a red background in the events table.
The Pass/Fail threshold LED, located on the front of the unit, will also
indicate the status (green for pass, red for fail).
You can also set the application to display pass/fail messages when the
Pass/Fail test is performed.
Test Default Minimum Maximum
Splice loss (dB) 1.000 0.015 5.000
Connector loss (dB) 1.000 0.015 5.000
Reflectance (dB) –40.00 –80.0 0.0
Fiber section attenuation (dB/km) 0.400 0.000 5.000
Span loss (dB) 45.000 0.000 45.000
Span length (km) 0.00 0.0000 300.0000
Span ORL (dB) 15.00 15.00 40.00
Przeglądanie stron 67
1 2 ... 63 64 65 66 67 68 69 70 71 72 73 ... 338 339

Komentarze do niniejszej Instrukcji

Brak uwag