Exfo FTB-700 for FTB-1 Instrukcja Użytkownika Strona 221

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 220
Analyzing Traces with the Bidirectional Analysis Application (Optional)
OTDR 213
Reanalyzing Traces and Regenerating the Bidirectional Measurement
Reanalyzing Traces and Regenerating the
Bidirectional Measurement
You can analyze the A->B and B->A traces and regenerate the
bidirectional measurement at any time. Reanalyzing a trace will:
re-create the events table if it was modified.
reset the span start to zero and the span end to end-of-fiber, unless you
have saved them (see Saving the Span-Start and Span-End Information
on page 244).
The table below shows what happens when you start the analysis,
depending on the direction that is currently selected.
Note: In the case of multiwavelength files, the analysis is performed for the
selected wavelength only.
If you prefer to focus your analysis on a specific fiber span, see Modifying
the Alignment of Unidirectional Traces on page 215.
Current direction
Bidirectional
measurement
A->B trace B->A trace
Bidirectional Regenerated Reanalyzed Reanalyzed
A->B Regenerated Reanalyzed Not modified
B->A Regenerated Not modified Reanalyzed
Przeglądanie stron 220
1 2 ... 216 217 218 219 220 221 222 223 224 225 226 ... 328 329

Komentarze do niniejszej Instrukcji

Brak uwag