SCPI Command Reference
Tunable Laser Source 125
Product-Specific Commands—Description
:TRIGger[1..n][:SEQuence]:INPut?
Description This query returns the current state for the input
trigger.
At *RST, the trigger is disabled.
Syntax :TRIGger[1..n][:SEQuence]:INPut?
Parameter(s) None
Response Syntax <TriggerMode>
Response(s) TriggerMode:
The response data syntax for <TriggerMode> is
defined as a <CHARACTER RESPONSE DATA>
element.
The <TriggerMode> response indicates the
current operating mode of the trigger.
DISABLED, trigger is disabled.
SSWEEP, trigger is enabled.
Example(s) TRIG:SEQ:INP SSW
TRIG:INP? returns SSWEEP (the trigger is
enabled)
See Also TRIGger[1..n][:SEQuence]:INPut
SOURce[1..n]:WAVelength:SWEep:MODE