Referencia de instrucciones SCPI
OTDR 453
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:POWer:STATe:
TIME?
Description This query returns a value indicating the duration
after which the visual fault locator (VFL) will
stop emitting light automatically (auto-off
feature).
*RST sets this value to 600 seconds.
Syntax :SOURce[1..n]:VFLocator:POWer:STATe:TIME?[
<wsp>MINimum|MAXimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MINimum|MAXimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Duration>