Caractéristiques techniques
344 FTB-610
PM-1 BUILT-IN POWER METER SPECIFICATIONS
a
Calibrated wavelengths (nm) 850, 1300, 1310, 1490, 1550, 1625, 1650
Optional CWDM calibrated
wavelengths (nm)
1270, 1290, 1310, 1330, 1350, 1370, 1390, 1410, 1430, 1450, 1470, 1490,
1510, 1530, 1550, 1570, 1590, 1610, 1383, 1625
Power range (dBm) 10 to —86 (InGaAs)
26 to —64 (GeX)
Uncertainty (%)
b
±5 % ± 3 pW (InGaAs)
±5 % ± 0.4 nW (GeX)
Display resolution (dB)
InGaAs
GeX
0.01 = max to —76 dBm
0.1 = —76 dBm to —86 dBm
1 = —86 dBm to min
0.01 = max to —54 dBm
0.1 = —50 dBm to —60 dBm
1 = —60 dBm to min
Automatic offset nulling range
c
Max power to —63 dBm for InGaAs
Max power to —40 dBm for GeX
Tone detection (Hz) 270/1000/2000
Notes
a. At 23 ºC ± 1 ºC, 1550 nm and FC connector. With modules in Idle mode. Battery-operated.
b. Up to 5 dBm.
c. For ±0.05 dB, from 18 ºC to 28 ºC.