Exfo PSO-100 Series Optical Sampling Oscilloscopes Instrukcja Użytkownika Strona 124

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 123
Measurement Definitions
116 PSO-100
Duty Cycle Distortion
Duty Cycle Distortion
The duty cycle distortion (DCD) is a measurement of the time separation of
the rising and falling edge of the eye at the center (50 %) level. The DCD is
defined in percentage of the bit period as:
In a "perfect" eye with no timing- and amplitude jitter, and a 50 % crossing,
the DCD is zero.
DCD 100%
T
rise,center
T
fall,center
bit period
------------------------------------------------------
=
50 %
Bit period
DCD time
Przeglądanie stron 123
1 2 ... 119 120 121 122 123 124 125 126 127 128 129 ... 139 140

Komentarze do niniejszej Instrukcji

Brak uwag