SCPI Command Reference
116 IQS-3150
Product-Specific Commands—Description
:OUTPut[1..n]:ALC[:STATe]?
Description This query indicates if the power tracking that
controls the output power level has been
activated or not. The power tracking is done via
the ALC (Automatic Leveling Control) loop.
At *RST, this value is set to off.
Syntax :OUTPut[1..n]:ALC[:STATe]?
Parameter(s) None
Response Syntax <State>
Response(s) State:
The response data syntax for <State> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <State> response corresponds to the state
of the ALC (Automatic Leveling Control) loop.
0: No power tracking will be performed.
1: Power tracking is activated.
See Also :OUTPut[1..n]:ALC[:STATe]
:CONTrol[1..n]:MODE
:OUTPut[1..n]:DTOlerance
:OUTPut[1..n]:POWer
:OUTPut[1..n]:RPOWer