Remote Control Commands
CD/PMD Analyzer Source 63
IEEE 488.2 Commands—Description
Response(s) RegisterValue:
The response data syntax for <RegisterValue> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <RegisterValue> ranges from 0 through 255.
The <RegisterValue> value expressed in base 2
(binary) represents the bit values of the Standard
Event Status Enable register. See below.
Example(s) *ESE? returns 133
where 133 = (bit PON, bit QYE and bit OPC)
See Also *ESE
*ESR?
*ESE?
Bit Weight Meaning
PON 128 Power ON Enable
N.U. 64 Not used
CMD 32 CoMmanD Error Enable
EXE 16 Execution Error Enable
DDE 8 Device Dependent Error Enable
QRY 4 QueRry Error Enable
N.U. 2 Not used
OPC 1 Operation Complete Enable