Exfo FastReporter 2 Instrukcja Użytkownika Strona 358

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 357
Description of Event Types
348 FastReporter 2
Merged Event
Merged Event
This symbol denotes an event combined with one or more other events. It
also indicates the total loss produced by the merged events following it in
the events table.
A Merged Event is composed of subevents. Only the Merged Event is
attributed a number in the events table, not the subevents composing
it, if they are displayed.
Reflective events may indicate the presence of connectors, mechanical
splices, or poor-quality fusion splices or cracks.
Non-reflective events may indicate the presence of splices, splitters or
bendings.
A reflectance value is specified for all merged events and indicates the
maximum reflectance for the merged event. A reflectance value is also
displayed for each reflective subevent composing the Merged Event.
Reflective events
Point A
Point B
Merged Reflective event position
Total loss ( dB)
Reflected power
(dB)
Distance
(km)